| Sonatest SITESCAN Narrow Band Amplified Ultrasonic
Flaw Detector SS250
Features:
This version includes all the software features of the SS150 combined with a narrow band amplifier and TCG for more demanding ultrasonic applications. The square wave ActiveEdge pulser automatically optimizes for each filter band selected.
Specifications SITESCAN Series
Test Range: 0 to 5 mm (0.2in) up to 0 to 10,000 (400in) at steel velocity. Variable in sequence, 10mm or 1mm
Velocity: 1000 to 9999 m/s
Probe Zero: 0 to 999,999 micro sec.
Delay: 0 to 10,000 mm at steek velocity in 0.05mm steps (0 to 400 inches) in 0.002 inch steps.
Gain: 0 to 110dB in 0.5, 1, 2, 6, 10, 14, and 20dB steps
Gates: Start & width adjustable over full range. Amplitude 0 to 100 %, 0.5% steps; visual & audible alarms. Two gates; Positive and negative triggering; Gate 2 has selectable 0.6 second delay on alarm.
Pulser Voltage: -200V peak amplitude; rise/fall time <10ns into 50 ohm.
Pulser Width: 30 to 250 ns linked to filter band.
Pulser Repetition Freq. Selectable 5 to 1000 Hz
Screen Update Rate: 50 or 60 Hz
Rectification: Full wave, postive or negative, RF
Frequency Range: Narrow bands at 1, 2.25, 5, 10 MHz, Broadband 1.5 to 15 MHz.
System Linearity: Vertical = 1 % Full Screen Height (FSH), Amplifier Accuracy +/- 0.1dB, Horizontal +/- 0.33% Full Screen Width (FSW)
Reject: 50 % suppressive. LED warning light
Units: mm/in/time micro seconds
Display: Colour Transflective TFT: Display area 111.4 x 83.5mm (4.39 x 3.29") 320 x 240 pixels. A-Scan Area 255 x 200 pixels (315 x 200 expanded), 8 colour options and variable brightmess.
A-Scan Memory: Maximum of 800 waveforms
Panel Memory: 100 stores for retaining calibrations
Thickness Logging: Storage for 8000 thckness readings
AGC: Sets selected echo to a user defined level (10-90%)
DAC: Defined by up to 10 points and digitally drawn on screen. DAC curves meet requirements of EN1714, JIS, and ASME selectable between -2, -6, -10, -12, and -14dB.
AWS: Automatic defect sizing in accordance with AWS D1.1 Structural Welding Code.
API: Automatic defect sizing in accordance with API 5UE
AVG/DGS: Automatic defect sizing using probe data
TCG: 40 dB dynamic range, 30dB per micro sec; up to 10 points for curve definition
Auto-Cal: Automatic calibration with 2 echoes
Reference Waveform: Recalled waveform can be shown in a different colour to live waveform for direct comparison
Display Freeze: Holds current waveform on screen
Peak Memory: For echo-dynamic pattern capture
Outputs: Full bi-directional serial interface. Composite video, PAL or NTSC compatibility, Analogue proportional outputs programmable to distance or ampliltlude of signal in the gate. Transmitter sync output front USB
Power Supply: Lithium Ion battery pack 14.4 V, 5.0 A-h, gives up to 16 hrs duration from a fully charged pack. Recharge time 3-4 hrs. Mains pack option.
Dimensions: 255 x 145 x 145mm (10 x 5.7 x 5.7")
Weight: 2.5 kg (5.5lbs) w/ Li-Ion cells.
|